Recall Z-2217-2024 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

NovaSignal Corp.

NovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2

DateJun 27, 2024 (initiated May 6, 2024)
StatusOpen, Classified
ReasonDue to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.
Root causeDevice Design
ActionOn May 6, 2024, NeuraSignal Inc. dba NovaSignal issued a "Urgent-Medical Device Recall" notification to affected consignees. Nova Signal asked consignees to take the following actions: 1. The NeuraSignal service and customer support team will schedule a time for the onsite field correction. 2. In the timeframe leading up to the scheduled onsite field correction, the system can be used as intended. 2a. Users of the device should refrain from accessing the enclosed cable connections during routine operation. This may lead to unintentional disconnection of the chassis ground cable from the AC
Quantity28 systems
DistributionWorldwide - US Nationwide distribution in the states of MA, TX, NY, MN, AZ, AL, CO, CA, OH, MO, FL, TN, NJ, GA, OK, KY, WA and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213279NovaGuide 2 Intelligent Ultrasound
Novasignal Corporation, Mar 2, 2022