Recall Z-2159-2024 ยท MNQ Stimulator, Hypoglossal Nerve, Implanted, Apnea

Inspire Medical Systems Inc.

Inspire Model 3028, IV Implantable Pulse Generator

DateJun 21, 2024 (initiated May 16, 2024)
StatusOpen, Classified
ReasonThere is a potential manufacturing defect which can lead to electrical leakage in the sensing circuit which causes a system malfunction and requires a revision surgery to replace the IPG,
Root causeProcess control
ActionInspire Medical systems issued an URGENT Medical Device Correction notice to the affected healthcare providers on 05/16/2024 via FedEx overnight. The notice explains the potential problem with the device and risk to the patient and recommends the following actions: We recommend the following actions: " Notify affected patients of this voluntary recall. " Schedule your patient for a monitoring visit to confirm their Inspire therapy is working as intended through waveform and impedance analysis. Monitor patients for reports of variation in stimulation, a lack of therapy effectiveness, or an
Quantity24 devices
DistributionWorldwide distribution - US Nationwide and the country of Germany.
FDA recordOpen on fda.gov