Recall Z-2160-2024 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

NxStage Medical Inc

NxStage PureFlow-B Solution, Premixed Dialysate for Hemodialysis, Device Catalog Numbers RFP-400, RFP-401, RFP-402, RFP-403, RFP-404, RFP-405, RFP-406, RFP-453, RFP-454, RFP-456

DateJun 20, 2024 (initiated May 14, 2024)
StatusOpen, Classified
ReasonThe smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE RECALL dated 5/14/24 was sent to customers. Action to be taken by Customers 1. Check all boxes in your Pureflow B solution inventory. If you find any boxes or individual bags of Pureflow B solution from the affected lots, separate them from your other inventory and do not use them. (Refer to Appendix B for how to locate lot number). 2. Follow the instructions on the attached reply form to acknowledge your receipt of this notice. 3. Contact NxStage Customer Service (acuteorders@freseniusmedicalcare.com) to arrange for the return of all affected product. We understa
Quantity1316936 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053286MODIFICATION TO: NXSTAGE PUREFLOW-B SOLUTION
Nxstage Medical, Inc., Dec 22, 2005