Recall Z-2029-2024 ยท HIH Hysteroscope (And Accessories)

Olympus Corporation of the Americas

OES 4000 Hysteroscope, Model/Catalog Number: A4674A

DateJun 6, 2024 (initiated Apr 18, 2024)
StatusOpen, Classified
ReasonThe required leakage test was not performed following of a re-work for a non-conformity. This could potentially result in liquid ingress into the eyepiece and cause a foggy image.
Root causeProcess change control
ActionOn May 9, 2024, URGENT: MEDICAL DEVICE REMOVAL letters were sent to the consignee. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Therefore, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Examine your inventory and identify the above listed device with serial number. 3. Please contact Customer Service at at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will arrange for the return of your device to Olympus, and we will replace this device a
Quantity1 unit
DistributionUS Nationwide distribution in the state of NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K897003OLYMPUS HYSTERORESECTOSCOPE