Recall Z-2135-2024 · OLO Orthopedic Stereotaxic Instrument

Blue Belt Technologies, Inc

CORI Starter Kit, Real Intelligence CORI, CORI STARTER KIT - UNIVERSAL.Intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures Product Numbers: (1) ROB10000 CORI Starter Kit (2) ROB10024 Real Intelligence CORI (3) ROB10197 CORI STARTER KIT - UNIVERSAL Note: the three product

DateJun 11, 2024 (initiated May 7, 2024)
StatusOpen, Classified
ReasonCORI Surgical System with Real Intelligence Software exhibits an increased Marker Registration Error (MRE) which may cause tracker arrays to flicker on screen of the CORI Surgical System during use, and may cause surgical delay
Root causeUnder Investigation by firm
ActionSmith & Nephew (Blue Belt wholly owned subsidiary of Smith & Nephew, Inc.) issued Urgent Medical Device Correction Notice C-2024-03 notification on May 7, 2024. Letter states reason for recall, health risk and action to take: Required Actions for user/hospital: 1. Please see attached IFU Addendum for troubleshooting tips and notes to enhance the consistency of visibility of the markers and tracking arrays. This IFU Addendum can be found at https://ifu.smith-nephew.com/ by searching IFU number 1000264245. 2. Complete Correction Notice Acknowledgement Form and submit to FieldActions@smith-nephe
Quantity916 units
DistributionNationwide Foreign: Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Puerto Rico, Saudi Arabia, Singapore, South Africa,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K240139Real Intelligence™ CORI™
Blue Belt Technologies, Inc., Mar 18, 2024