ClearanceBrief
Recall Z-2010-2024 ยท DSY Prosthesis, Vascular Graft, Of 6mm And Greater Diameter

Atrium Medical Corporation

FLIXENE, 7X50, 1GDS, STR. Single-Ended Slider GDS. Vascular graft

DateJun 4, 2024 (initiated May 3, 2024)
StatusOpen, Classified
ReasonComplaints of reported separation of the Slider GDS Swivel Rod from the Swivel Core and of a notable gap between the two pieces.
Root causeEquipment maintenance
ActionConsignees were sent an URGENT MEDICAL DEVICE- REMOVAL notification by mail, dated 5/3/24. Consignees are instructed to examine their inventory immediately to determine if any recalled product is in their possession. Recalled product possesses the REF number on device labeling and has an expiration date prior to 24FEB2027. All consignees are asked to complete and return the provided response form by email to GDSgrafts2024@getinge.com or by fax to 1-866-560-4062. Consignees that are distributors are asked to forward the provided notification to their customers. Consignees are to post a copy of
Quantity53,308 units (US: 11, 236; OUS: 42,072)
DistributionWorldwide distribution - US Nationwide including Puerto Rico, and the countries of AE, AR, AT, AU, BE, BG, BO, BR, CA, CH, CL, CN, CO, CY, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GB, GR, HK, ID, IE, IL, IQ, IT, JO, JP, KR, KW, LT, LU, LV, ME, MN, MU, MX, MY, NI, NL, NO, NZ, OM, PA, PE, PH, PK, PL, PT, P
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060124ATRIUM FLIXENE GRAFT
Atrium Medical Corp., Mar 30, 2006
K071923PTFE FLIXENE GRADUATED WALL GRAFT
Atrium Medical Corp., Aug 14, 2007