ClearanceBrief
Recall Z-2172-2024 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

GE Healthcare Finland Oy

CARESCAPE B850 Model Numbers: a) MBC303, Model Numbers: 1) 2078633-001, 2) 2078633-010, 3) 2078633-012, 4) 2078633-017, 5) 2078633-028, 6) 2095802-001-54069330, 7) 2095802-001-554922, 8) 2095802-001-554923, 9) 2095802-001-554924, 10) 2095802-001-554925, 11) 2095802-001-554926, 12) 2095802-001-554927, 13) 2095802-001-554928, 14) 2095802-001-555120, 15) 2095802-001-556223, 16) 20

DateJun 21, 2024 (initiated Apr 5, 2024)
StatusOpen, Classified
ReasonGE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
Root causeSoftware design
ActionGE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 04/05/2024 via *****. The notice explained the problem with the device, potential risk, and requested the following actions to be taken by Customer/User: 1. You can continue to use your monitor if the CPU timekeeper battery has not been replaced. If the monitor has had a CPU timekeeper battery replaced or if you are unsure whether the CPU timekeeper battery was replaced, reinstall the host software using the software media that is included as part of the equipment. 2. In the event the monitor does not start u
Quantity25268 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191149CARESCAPE B650
GE Healthcare Finland Oy, Jan 21, 2020
K213181CARESCAPE B650, E-musb
GE Healthcare Finland Oy, Apr 13, 2022