Recall Z-1948-2024 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America Llc

HA FlexTrak-Patient transport functionality to transport the patient from the preparation area to the MR system, Product Number: 989710006411

DateJun 4, 2024 (initiated May 3, 2024)
StatusOpen, Classified
ReasonOil may leak from the HA FlexTrack Trolley onto the floor when pressing the hydraulic pedal, resulting in a potential slipping/falling hazard
Root causeUnder Investigation by firm
ActionPhilips NA issued URGENT Medical Device Correction letter to US consigneess dated May 3, 2024 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the US. Letter states reason for recall, health risk and action to take: A. If you see hydraulic fluid (oil) leakage: " First remove fluid from the floor using a paper towel. Then remove any remaining oil film from the floor using a cleaning agent (e.g. soap and water) and paper towel. " Contact your local Philips service representative. B. You may continue to use your HA FlexTrak Trolley in accordance w
Quantity1348 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Fiji, Finland, France, French Guiana, Germany, Guadeloupe, Honduras, Hong Kong, India, Indonesia, Ireland, Italy, J
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213583Achieva, Ingenia, Ingenia CX, Ingenia Elition and Ingenia Ambition MR Systems
Philips Medical Systems Nederland B.V., May 16, 2022