ClearanceBrief
Recall Z-1868-2024 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Holdings, Inc.

2008T BlueStar Hemodialysis Machine, Model 191126 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.

DateMay 17, 2024 (initiated Apr 17, 2024)
StatusOpen, Classified
ReasonAn increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
Root causeDevice Design
ActionOn April 17, 2024 IMPORTANT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer actions: As per the 2008T Operators Manual P/N 490122 Rev AA: Blood Pump Rotors should be inspected for proper operation, as stated in Chapter 2: Daily preparation for treatment. If you experience a dislodged guide post sleeve, bent or loose tubing guide posts, or roller cannot move freely on your current F40015481 Rev A blood pump rotor, please contact FMCRTG Technical Support Services at 800-405-1321 for a replacement Blood Pump Rotor (Part Number, F40015481, Rev C) . Have Rotor Lot Number availab
Quantity12,292 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1739722008T BlueStar Hemodialysis Machine
Fresenius Medical Care Renal Therapies Group, LLC, May 10, 2018
K2229522008T BlueStar Hemodailysis Machine
Fresenius Medical Care Renal Therapies Group, LLC, Oct 27, 2022
K2311252008T BlueStar Hemodialysis Machine
Fresenius Medical Care Renal Therapies Group, LLC, Dec 6, 2023