ClearanceBrief
Recall Z-2236-2024 ยท KWY Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented

Exactech, Inc.

Acumatch L-Series BIPOLAR LINER, 22mm (BIPOLAR ENDOPROSTHESIS-Hip), Item Numbers: a) 100-22-19, SZ. L; b) 100-22-20, SZ. M; c) 100-22-21, SZ. N; d) 100-22-22, SZ. P; e) 100-22-23, SZ. R; f) 100-22-24, SZ. S; g) 100-22-25, SZ. T

DateJun 28, 2024 (initiated Apr 26, 2024)
StatusOpen, Classified
ReasonThe AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), Nylon, and EVOH, or 2) LDPE and Nylon without EVOH.
Root causeOther
ActionExactech issued an UREGNT VOLUNTARY MEDICAL DEVICE RECALL notice to its sales representatives on 04/26/2024 via email. The notice explained the issue, clinical impact (risk), and requested the following: Actions to be Taken: " Review this notification thoroughly. " Immediately discontinue use and quarantine any affected product. " Distribute this Recall Letter to all the surgeons who have implanted the affected devices. " Send all affected product back to Exactech as outlined in the attached Recall Confirmation Form .
Quantity589 units
DistributionWorldwide - US distribution in Puerto Rico and the countries of Great Britain, Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013211MODIFICATION TO: ACUMATCH L-SERIES BIPOLAR ENDOPROSTHESIS
Exactech, Inc., Dec 5, 2001