Recall Z-2026-2024 ยท OWB Interventional Fluoroscopic X-Ray System

Medtronic Navigation, Inc.-Littleton

O-arm O2 Imaging System-Mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients. Product Number: BI70002000

DateJun 6, 2024 (initiated Apr 26, 2024)
StatusOpen, Classified
ReasonPotential for an electrical component of the O-arm" O2 Imaging System to intermittently shut off,failure can cause the potential for surgical delay, additional imaging, cancellation and rescheduling of surgery, or completion of surgery without the benefit of image guidance.
Root causeDevice Design
ActionMedtronic issued Urgent Medical Device Correction Letter on 4/26/24 to US end-users hand delivered via Field Service representative. OUS communications were initiated with affected consignees on April 17, 2024. Letter states reason for recall, health risk and action to take: A Medtronic Field Service Engineer will contact your facility to schedule time to service your system within 90 days. Required Actions: 1. If your system is actively in use, you may continue to use it pending service by Medtronic. If your system exhibits the issue, it will automatically reset after beeping. If the frequen
Quantity20 units
DistributionWorldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, MN, PR, TX, WI and the countries of Bangladesh, Canada, Costa Rica, India, Mexico, Panama, Philippines.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151000O-Arm O2 Imaging System
Medtronic, Inc., Aug 6, 2015