ClearanceBrief
Recall Z-1916-2024 ยท HXX Screwdriver

Synthes (USA) Products LLC

STARDRIVE SCRWDRVR SHFT/T4 50MM/SELF-RETAINING/HXC-Intended for tightening, loosening, or removing of screw implants during orthopaedic surgery. Part Number: 03.130.010

DateMay 28, 2024 (initiated Apr 25, 2024)
StatusOpen, Classified
ReasonStardrive Screwdrivers manufactured without a 5-degree relief cut around the tip and will not engage with their respective 1.3mm and 1.5mm screws, may result in surgical delay
Root causeProcess change control
ActionDePuy Synthes issued Urgent Medical Device Recall (Removal) Letter on 4/26/24. Letter states reason for recall, health risk and action to take: Examine your inventory to determine if you have the subject product lots and quarantine them immediately. DO NOT USE THE SUBJECT PRODUCTS. 2. Contact your DePuy Synthes Sales Consultant or contact the customer support services at ra-dpyuscsspecial@ its.jnj.com to coordinate the return/credits of the subject product lots. 3. Review, complete, sign, and return the attached Business Response Form (page 3 of this letter) to OneMDField- Actions@its.jnj.com
Quantity261 units
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall