Recall Z-1940-2024 ยท MOF Guide, Wire, Catheter, Neurovasculature
Stryker Neurovascular
The Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip
| Date | May 31, 2024 (initiated Apr 18, 2024) |
| Status | Open, Classified |
| Reason | Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with im |
| Root cause | Component design/selection |
| Action | On 4/29/24, recall notices were mailed to Supply Chain Management/ Recall Coordinator/ Inventory Manager customers who were asked to do the following: 1) When using impacted guidewires, please do not backload the guidewire, follow the Instructions for Use (IFU) provided with the device, and do not use the guidewire if you see any visible signs of coating damage. 2) Circulate this customer notification internally to all interested/affected parties. 3) Inform firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that the firm can inf |
| Quantity | 41,074 |
| Distribution | Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENT |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K202522 | Synchro SELECT Guidewire Stryker Neurovascular, Sep 25, 2020 |