Recall Z-1933-2024 ยท QVA Fludeoxyglucose F18-Guided Radiation Therapy System

Reflexion Medical, Inc.

RefleXion X1 Radiotherapy System Model RXM1000 Software version 2.1.19-3 Radiation therapy system is a device that combines the functionality of an emission computed tomography detection system and a linear accelerator.

DateMay 30, 2024 (initiated Apr 16, 2024)
StatusOpen, Classified
ReasonA potential dose error exists for patients treated with an out of session SCINTIX partial fraction.
Root causeSoftware design
ActionOn 04/29/2024, the firm sent an "Urgent: Medical Device Recall Safety Notification" via email to customer informing them that during internal dosimetric testing of SCINTIX OOS partials in periodic tracking mode, it was observed that the delivered dose error of the full fraction can exceed 5% with respect to the upper bound of one or more static volumes within the DVH in the plan (non-tracked, non-target volumes in the patient frame of reference, i.e., the planning CT). This was determined to be the result of a software defect associated with completing interrupted. Customers are instructed to
Quantity7 systems
DistributionU.S.: CA, CT, NJ, OR, PA and TX No O.U.S.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.