Recall Z-1930-2024 ยท BRY Cabinet, Table And Tray, Anesthesia
CareFusion 303, Inc.
BD Pyxis MedStation 4000 7-Drawer Auxiliary Tower, REF: 306
| Date | May 30, 2024 (initiated Apr 15, 2024) |
| Status | Open, Classified |
| Reason | Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire. |
| Root cause | Component design/selection |
| Action | On May 2, 2024, CareFusion, a subsidiary of Becton, Dickinson and company issued a "Urgent Medical Device Correction", to affected consignees. On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions: 1. Affected products can continue to be used per existing Manuals and Guides. 2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification. 3. Review the customer letter attachment which |
| Quantity | 35,029 units |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.