ClearanceBrief
Recall Z-1929-2024 ยท BRY Cabinet, Table And Tray, Anesthesia

CareFusion 303, Inc.

BD Pyxis" MedStation ES 7-Drawer Auxiliary Tower, REF: 324

DateMay 30, 2024 (initiated Apr 15, 2024)
StatusOpen, Classified
ReasonPotential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
Root causeComponent design/selection
ActionOn May 2, 2024, CareFusion, a subsidiary of Becton, Dickinson and company issued a "Urgent Medical Device Correction", to affected consignees. On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions: 1. Affected products can continue to be used per existing Manuals and Guides. 2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification. 3. Review the customer letter attachment which
Quantity87,302 units
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.