Recall Z-2264-2024 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter, Inc.

DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic

DateJul 3, 2024 (initiated Apr 18, 2024)
StatusOpen, Classified
ReasonBeckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids, by correction. The reason for the recall is: Beckman Coulter has determined that the DxI 9000 Access Immunoassay analyzer accepts calibrator values scanned from multiple calibrator cards when an operator configures a new calibrator lot. The analyzer accepts the calibrator lot values from any scanned calibrator card, regardless of the assay test code or cali
Root causeSoftware design
ActionBeckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 04/18/2024 via USPS first class mail. The notice explained the issue, impact, and requested the following actions be taken: "x Confirm that the DxI 9000 Access Immunoassay Analyzer test name printed on the calibrator card matches the calibrator material name before you configure the calibrator lot. x Scan each individual calibrator card and complete each step displayed on the DxI 9000 Access Immunoassay Analyzer user interface before you scan the next calibrator card. x Review the expected values column on th
Quantity116 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221225Access TSH (3rd IS) Assay, DxI 9000 Access Immunoassay Analyzer
Beckman Coulter, Inc., Nov 10, 2022