ClearanceBrief
Recall Z-1887-2024 ยท JZO System, Test, Thyroid Autoantibody

Siemens Healthcare Diagnostics, Inc.

IMMULITE 2000 Anti-TG Ab Catalog # L2KTG6 (600 Tests), SMN 10381655, System, Test, Thyroid Autoantibody.

DateMay 23, 2024 (initiated Apr 23, 2024)
StatusOpen, Classified
ReasonThe potential for falsely elevated patient sample results when using impacted lots (785, 790, 791, and 793) of IMMULITE 2000 Anti-TG Ab.
Root causeUnder Investigation by firm
ActionOn April23, 2024, Urgent Medical Device Correction letters were sent to customers. Customer Actions: 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2. Perform the following: Discontinue use of and discard the kit lots listed in the Products Table. You may request no-charge replacement product from your local Siemens Healthineers or distributor office. Please review your inventory of these products to determine your laboratory s replacement needs and to provide information to
Quantity488 units (33 US, 455 OUS)
DistributionWorldwide - US Nationwide distribution including in the states of AL, CA, DC, FL, HI, ID, IN, KS, KY MA, MI, MN, MO, NC, NJ, NY, OH, OR, PA, PR, TX, WA WI and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, Croatia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991094IMMULITE 2000 ANTI-TG AB, MODEL L2KTG2, L2KTG6
Diagnostic Products Corp., May 13, 1999