ClearanceBrief
Recall Z-1830-2024 ยท LNH System, Nuclear Magnetic Resonance Imaging

Philips North America Llc

Ingenia 3.0T with Evolution Upgrade 3.0T - - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782117

DateMay 15, 2024 (initiated Apr 22, 2024)
StatusOpen, Classified
ReasonPhilips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation
Root causeSoftware design
ActionPhilips issued URGENT Medical Device Correction letter on 4/22/24 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: a. Customers can continue to use the identified system(s) in accordance with the intended use pertaining to setting patient ventilation. (Reference IFU, including below information). {If your Ingenia 1.5T or Ingenia 3.0T system(s) has been upgraded to Ingenia Evolution configuration and running R5.7 software, it is impacted by this issue. (Note: customer wil
Quantity19 units
DistributionWorldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the countries of Austria, Croatia, Finland, France, Germany, Italy, Lebanon, Netherlands, Norway, Sweden, Switzerland, Thailand, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems
Philips Medical Systems Nederland B.V., Apr 10, 2020