Recall Z-1829-2024 ยท LNH System, Nuclear Magnetic Resonance Imaging
Philips North America Llc
Ingenia 1.5T with Evolution Upgrade 1.5T - Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device REF: 782116
| Date | May 15, 2024 (initiated Apr 22, 2024) |
| Status | Open, Classified |
| Reason | Philips Ingenia Evolution systems with software R5.7.1 may not provide warning for low ventilation setting which may lead to increase in body temperature or burning sensation |
| Root cause | Software design |
| Action | Philips issued URGENT Medical Device Correction letter on 4/22/24 via FedEx. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Letter states reason for recall, health risk and action to take: a. Customers can continue to use the identified system(s) in accordance with the intended use pertaining to setting patient ventilation. (Reference IFU, including below information). {If your Ingenia 1.5T or Ingenia 3.0T system(s) has been upgraded to Ingenia Evolution configuration and running R5.7 software, it is impacted by this issue. (Note: customer wil |
| Quantity | 32 units |
| Distribution | Worldwide - US Nationwide distribution in the states of CA, KS, MS, NH and the countries of Austria, Croatia, Finland, France, Germany, Italy, Lebanon, Netherlands, Norway, Sweden, Switzerland, Thailand, United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K193215 | Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems Philips Medical Systems Nederland B.V., Apr 10, 2020 |