Recall Z-1817-2024 ยท MNS Ventilator, Continuous, Non-Life-Supporting

Philips Respironics, Inc.

OmniLab Advanced + Ventilator Product Numbers 1111122 1111123 1111124 1111125 1111126 1111127 1111138 1111141 1111142 KR1111127 R1111122 R1111123 R1111124 U1111122 U1111124

DateMay 23, 2024 (initiated Apr 1, 2024)
StatusOpen, Classified
ReasonDevice may experience an interruption or loss of therapy in case of a Ventilator Inoperative alarm. This may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death.
Root causeDevice Design
ActionFirm notified customers through mailed letter "URGENT Medical Device Recall" on April 1, 2024. The letter described the product, problem and actions to be taken. The firm reminded customers to refer to the user manual and to perform a clinical assessment prior to placing a patient on the ventilator to ensure that: 1. The device is appropriately set for patient requirements. 2. Alternate ventilation equipment is available. 3. Alternative monitoring is used where appropriate. Depending on the level of ventilatory support required, customers were advised to place patients on an alternate sour
Quantity9,388 US; 2,664 ROW
DistributionUS Nationwide distribution and OUS International distribution to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090539BIPAP AUTOSV ADVANCED
Respironics Inc., Sleep & Home Respiratory Group, Oct 30, 2009
K113068REMSTAR AUTO A-FLEX HT
Respironics, Inc., Dec 16, 2011
K121623BIPAP A 40 VENTILATORY SUPPORT SYSTEM
Respironics, Inc., Dec 14, 2012