Recall Z-1944-2024 ยท FPA Set, Administration, Intravascular

BD SWITZERLAND SARL

Alaris Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite Y Sites REF 10013072 Tubing set used for delivery of fluids, medication, blood and blood products.

DateMay 31, 2024 (initiated Apr 18, 2024)
StatusOpen, Classified
ReasonDue to infusion set tubing drip chamber becoming detached from the tubing potentially causing delays in treatment, or unintentionally exposing users to harsh infusates.
Root causeProcess control
ActionOn 04/18/2024, the firm sent an URGENT: MEDICAL DEVICE PRODUCT RECALL" Letter via FedEx Express Saver to customer informing them that BD has received customer reports of the Alaris" Pump Infusion Set Low Sorbing Tubing (PE Lined) drip chamber becoming detached from the tubing for the catalog number and lot number referenced above. BD has confirmed that this issue is isolated to catalog number, 10013072 only and a single lot (no. 22115368) where mis-assembly of the drip chamber to the tubing occurred. Customers are instructed to: 1. Ensure the contents of this Product Recall communication are
Quantity11.300 sets
DistributionWorldwide - US Nationwide distribution in the states of AR, CA, CO, IL, KS, MA, ME, NJ, NY, PA, SD, WA and WI. The country of Canada.