Recall Z-1827-2024 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips North America Llc

Patient Information Center iX and Patient Information Center iX Expand, Software Version 4.x.

DateMay 15, 2024 (initiated Apr 19, 2024)
StatusOpen, Classified
ReasonEvent Catalog information does not save when copied and transferred from one unit to another.
Root causeSoftware design
ActionConsignees were sent an "URGENT Medical Device Correction," notification, dated 4/12/24. Consignees are instructed in the notification to review in Events Monitor and confirm the expected category is assigned for the reviewed event. If events are not assigned to the correct category consignees are not to depend on event notification for patient monitoring. The provided recall notification should be provided to all users of affected devices and to those the device was further distributed, for their awareness. Philips will contact consignees to schedule a visit from a Philips Field Service Engin
Quantity358 units (US: 45; OUS: 313)
DistributionWorldwide - US Nationwide distribution in the states of CA, FL, GA, IA, MA, MI, MN, MO, NJ, NM, NY, OH, PA, TN, TX, WA, WV, & DC. The countries of AE, AT, AU, BE, BR, CA, CH, CZ, DE, DK, ES, FI, FR, GB, IE, IL, IN, IT, JP, KR, LT, NL, NO, OM, PL, PT, RO, SE, SG, SK, TH, & ZA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K153702M3290B Philips IntelliVue Information Center iX
Philips Medical Systems, Jun 13, 2016