ClearanceBrief
Recall Z-1825-2024 ยท GWM Device, Monitoring, Intracranial Pressure

Sophysa S.A.

The Sophysa Pressio Ventricular Intracranial Pressure Monitoring Kit monitoring is intended for monitoring intracranial ventricular pressure with tunneling (PSO-VT) indicated for use in intraventricular pressure monitoring and cerebrospinal fluid drainage application. Model/Catalog Number: PSO-VT

DateMay 15, 2024 (initiated Apr 9, 2024)
StatusOpen, Classified
ReasonAn increase in claims related to leakage of cerebrospinal fluid (CSF) from the proximal end of the catheter. In the event of a leak going undetected during implantation, it could pose an infectious risk to the patient. The cause of the leaks is indicated to be manufacturing error involving perforation of the inner channel of the catheter during stylet insertion.
Root causeProcess change control
ActionSophysa issued Urgent Product Recall Letter dated 4/9/24 via email. Letter states reason for recall, health risk and action to take: 1- Please stop using the listed products immediately and remove all affected units from your inventory, 2- Please inform the healthcare professionals at your facility who use these products, 3- Isolate the units concerned in a safe place for return to Sophysa 4- Please complete the enclosed Acknowledgement Form, even if you are no longer in possession of the products to be returned. And return it to Sophysa contact@sophysa.us no later than April 15, 2024. 5- If y
Quantity3487 units OUS
DistributionWorldwide - US Nationwide distribution in the states of AZ, NY, TX and the countries of ARGENTINA, CHINA, CZECH REPUBLIC, FRANCE, GREECE, HUNGARY, JAPAN, PAKISTAN, Portugal, ROMANIA, SOUTH AFRICA, SWEDEN, VIETNAM.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062584PRESSIO ICP MONITORING SYSTEM
Sophysa SA, Jul 5, 2007
K162108Pressio 2 ICP Monitoring System
Sophysa SA, Apr 17, 2017