Recall Z-1909-2024 ยท DQX Wire, Guide, Catheter

Stryker Neurovascular

Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Catalog: M00313310 SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310 SYNCHRO .010 WIRE SOFT STRAIGHT 300cm, Catalog: M00316330

DateMay 28, 2024 (initiated Apr 23, 2024)
StatusOpen, Classified
ReasonNeurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Root causeProcess design
ActionOn 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes,
Quantity2859
DistributionWorldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Aus
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002907PVS 1300 NEURO GUIDE WIRE
Precision Vascular Systems, Inc., Mar 8, 2001
K023700PVS 1600 SYNCHRO 0.010 NEURO GUIDE WIRE
Precision Vascular Systems, Inc., Dec 4, 2002