Recall Z-1906-2024 ยท QJP Catheter, Percutaneous, Neurovasculature

Stryker Neurovascular

Trevo Trak 21 Microcatheter, Catalog: 90338

DateMay 28, 2024 (initiated Apr 23, 2024)
StatusOpen, Classified
ReasonNeurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Root causeProcess design
ActionOn 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes,
Quantity899
DistributionWorldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Aus
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211594Trevo Trak 21 Microcatheter
Stryker Neurovascular, Nov 26, 2021