Recall Z-1873-2024 ยท HBG Drills, Burrs, Trephines & Accessories (Manual)

Integra LifeSciences Corp.

Integra Cranial access kit (without Prep solutions) Bit and Guard- Used during a ventriculostomy procedure. Catalog Number: INS5HND

DateMay 20, 2024 (initiated Apr 17, 2024)
StatusOpen, Classified
ReasonPotential holes and tears in the sterile packaging (header bag) of the Cranial Access Kit may compromise the sterility of the device.
Root causeProcess control
ActionIntegra LifeSciences issued First Notification: Urgent: Voluntary Medical Device Recall Letter delivered by courier service, facsimile or email between April 17th and April 18th, 2024. Letter states reason for recall, health risk and action to take: 1. If you do have the affected product listed in Table 1, remove the product from further distribution. 2. Complete the attached Acknowledgment Form. 3. If you do have affected product, check the box I do have affected product. Record the total quantity of affected product you have. 4. If you do not have affected product, check the box, I do no
Quantity1,822 units
DistributionWorldwide distribution - US Nationwide and the countries of CA, BE, CL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961113CNS INTERCHANGEABLE BIT CRANIAL DRILL
Clinical Neuro Systems, LLC, Jun 5, 1996