Recall Z-1831-2024 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Maquet Medical Systems USA

CARDIOHELP-i System. Blood oxygenation and carbon dioxide removal system.

DateMay 16, 2024 (initiated Apr 15, 2024)
StatusOpen, Classified
ReasonThe firm identified that the measured patient leakage current in some devices significantly exceeded expected levels.
Root causeUnder Investigation by firm
ActionConsignees received an "URGENT MEDICAL DEVICE - Removal" email, dated 4/15/2024, notifying them of this recall event. Consignees are asked to examine their inventory to identify if they possess any affected devices and all consignees are asked to return the provided Response Form via email to identify where affected devices are located. Consignees that report having any affected devices will be contacted by Getinge to arrange for return of the affected device to upgrade the system's sensor panel and perform testing to ensure that leakage current is within specifications. Technical Support can
Quantity148 units (US: 5; OUS: 143)
DistributionUS Nationwide distribution in the states of NJ & TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133598CARDIOHELP SYSTEM
Maquet Cardiopulmonary, AG, May 21, 2014