Recall Z-1883-2024 · PHX Shoulder Prosthesis, Reverse Configuration

Tornier S.A.S.

Stryker Blueprint Software, Catalog #BPUE001.

DateMay 22, 2024 (initiated Apr 16, 2024)
StatusOpen, Classified
ReasonThe software bug allows for case planning with anatomic glenoid Perform / Perform Augmented implants and anatomic humeral Tornier Flex implant configurations that are incompatible and do not have regulatory approval.
Root causeSoftware change control
ActionThe firm issued letters via email on 4/16/2024. The email provided the issue and risks and said that for additional information regarding risks, exact incompatible configuration details, and steps for you to take, please refer to the attached document. The email listed a requested action, which was to review the case information provided on the business acknowledgment form, confirm the final implants used in this surgery, and complete and sign the response form on Page 4 at their earliest convenience. The attached Urgent Medical Device Recall letter was dated 16-April-2024 and provided the
DistributionUS Nationwide distribution in the states of FL, KY, NJ, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211359Tornier Perform™ Patient-Matched Primary Reversed Glenoid and BLUEPRINT™ Patient Specific Instrumentation
Tornier, Inc., Nov 12, 2021