ClearanceBrief
Recall Z-1884-2024 · IYN System, Imaging, Pulsed Doppler, Ultrasonic

FUJIFILM Healthcare Americas Corporation

Noblus AC Adapter used with the Noblus Ultrasound Imaging System Model : AHM250PS24

DateMay 22, 2024 (initiated Apr 11, 2024)
StatusOpen, Classified
ReasonAC adapters of Noblus can heat up, and cause smoke in the worst case, when an excessive mechanical stress is applied to the socket for power cable (frequent insertion/removal of the power cable).
Root causeDevice Design
ActionFujiFilm issued Urgent Medical Device Recall Letter via email on 4/11/24. Letter states reason for recall, health risk and action to take: Please complete and return the Field Action Verification Form attached on page 3 of this communication. Responding with the information requested is essential for ensuring appropriate action is taken. Please accept the replacement of AC adapter that is implemented the countermeasure. Please contact our field action coordinator, Kotei Aoki, via email at HCUSRegulatoryAffairs@fujifilm.com, if you have any further questions regarding this field action.
Quantity418 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160559NOBLUS™ Ultrasound Diagnostic System
Hitachi Medical Corporation, Oct 18, 2016