Recall Z-1902-2024 ยท POL Neurovascular Mechanical Thrombectomy Device For Acute Ischemic Stroke Treatment

Stryker Neurovascular

Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XP ProVue Retriever 4 x 30, Catalog: 90185 Trevo XP ProVue Retriever 6 x 25, Catalog: 90186 Kits: TREVO XP 3X20+PRO14 2-PACK, Catalog: 80051 TREVO XP 4X20+PRO18 2-PACK, Catalog: 80052 TREVO XP 4X30+PRO18 2-PACK, Catalog: 93067 TREVO XP 6X25+XT27 2-PACK, Catalog: 93068 TREVO XP 3X20+PRO14+CAT6

DateMay 28, 2024 (initiated Apr 23, 2024)
StatusOpen, Classified
ReasonNeurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Root causeProcess design
ActionOn 4/23/24, recall notices were mailed to Supply Chain Management, Recall Coordinators, and Inventory Managers who were asked to do the following: 1) Cease further distribution or use of all affected devices. 2) Segregate the affected units in a secure location for return to the firm. 3) Circulate this Recall-Removal notice internally to all interested/affected parties. 4) Maintain awareness of this communication internally until all required actions have been completed within your facility. 5) Inform the firm if any of the subject devices have been distributed to other organizations. If yes,
Quantity461
DistributionWorldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Aus
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173352Trevo Pro Vue Retriever and Trevo XP Pro Vue Retriever (Trevo Retriever)
Concentric Medical, Inc., Feb 15, 2018