Recall Z-1819-2024 ยท JIS Calibrator, Primary

QUIDELORTHO

VITROS Immunodiagnostic Products Free T3 Calibrators-IVD are for use with Free T3 Reagent Pack for the quantitative measurement of free triiodothyronine (FT3) in human serum and plasma (EDTA or heparin). Product Code: 111 2820

DateMay 14, 2024 (initiated Apr 9, 2024)
StatusOpen, Classified
ReasonFree T3 Calibrators used with T3 Reagent Packs may produce positively biased results for patient and quality control samples. The positive bias may cause erroneous patient results or a delay in testing that may impact patient management, clinical assessment, and the results of other thyroid function tests.
Root causeUnder Investigation by firm
ActionQuidelOrtho issued Important Product Correction Notification issued April 9, 2024. Letter states reason for recall, health risk and action to take: Discontinue using, render unusable, and discard affected lots of VITROS Free T3 Reagent Pack (and associated calibrators). " Complete the enclosed Confirmation of Receipt form no later than April 17, 2024. Upon receipt of your completed form, QuidelOrtho will provide credit for, or replacement of, discarded product. " Save this notification with your User Documentation or post this notification by each VITROS ECi/ECiQ/3600/5600/XT 7600 System until
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore , Spain, Sweden, The Netherlands, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K994145VITROS IMMUNODIAGNOSTIC PRODUCTS FREE T3 REAGENT PACK, VITROS IMMUNODIAGNOSTIC PRODUCTS FREE T3 CALIBRATORS
Ortho-Clinical Diagnostics, Inc., Dec 22, 1999