ClearanceBrief
Recall Z-1708-2024 ยท EIH Powder, Porcelain

IVOCLAR VIVADENT AG

IPS e.max ZirCAD CER/in. Prime C2 C17/5- A zirconium oxide block for the fabrication of fixed, full-contour restorations in anterior and posterior teeth. Article Number: 758443

DateApr 30, 2024 (initiated Apr 5, 2024)
StatusOpen, Classified
ReasonDefective blocks may show cracks after sintering If cracks are unnoticed, restorations may fracture in the patient's mouth and require restoration
Root causeUnder Investigation by firm
ActionIvokar issued Urgent Use and Safety Notice (03-2024-003-R) Letter to Distributors/End users via Fed'X or Email on 4/05/24. Letter states reason for recall, heath risk and action to take: 1. Identification of the IPS e.max ZirCAD Prime CEREC/inLab, C17 Block batches affected by the recall. 2. Separate the blocks so that they can no longer be used for the fabrication of dental restorations. 3. Contact the Customer Service of the distributor from whom you purchased the affected IPS e.max ZirCAD Prime CEREC/inLab, C17 Blocks to announce the return of the devices. Please note on the delivery bill
Quantity2472 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051705IPS E.MAX CAD/IPS E.MAX ZIRCAD
Ivoclar Vivadent, Inc., Oct 14, 2005