ClearanceBrief
Recall Z-1747-2024 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation 10.1.0.613 and 10.1.1.54. Radiation Therapy Treatment Planning System.

DateMay 1, 2024 (initiated Mar 28, 2024)
StatusOpen, Classified
ReasonPotential for reported SSD to be too high.
Root causeSoftware design
ActionA Field Safety Notice for Medical Device Correction #133261, dated 3/28/24, was sent to consignees via email. Consignees were provided the instructions to use the Measure Tool to measure the distance in the patient view to make sure the reported SSD value is correct if it is to be used for patient setup or other critical tasks. If the SSD value is found to be incorrect, small changes to the isocenter, beam angles or the affected ROI's geometry can be made until the correct value is achieved. Consignees are to inform planning staff and all users of the workarounds provided in the Field Safety N
Quantity18 units
DistributionWorldwide distribution - US Nationwide and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, France, Germany, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Korea, Malaysia, Morocco, Netherlands, New Zeal
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K210645RayStation 10.1
RaySearch Laboratories AB (PUBL), Jun 29, 2021