ClearanceBrief
Recall Z-1791-2024 ยท HWC Screw, Fixation, Bone

TriMed Inc.

TriMed, Threaded IM Nail 3.6mm, 60mm, TI, Ref: IMN3.6-60 T, Non Sterile, MD, Rx Only TriMed, Threaded IM Nail 3.6mm, 65mm, TI, Ref: IMN3.6-65 T, Non Sterile, MD, Rx Only

DateMay 10, 2024 (initiated Mar 19, 2024)
StatusOpen, Classified
ReasonNail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
Root causeProcess control
ActionOn 4/3/24, TriMed Inc., issued a "Urgent: Medical Device Recall", to affected consignees via E-Mail. TriMed ask consignees to take the following actions: " If you have the affected product, discontinue further use and quarantine products prior to return. " If a successful implantation was completed no further action is necessary. Only standard monitoring of the patient is required after successful implantation. " Please share this notice with all personnel that may utilize the affected products. Please share this notification to any locations where the affected product may have been t
Quantity142 units
DistributionUS Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, OH, OK, PA, TN, TX, UT, VA, WA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K234040TriMed Threaded Intramedullary Nail System
TriMed, Inc., Jan 18, 2024