Recall Z-1935-2024 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

CORENTEC CO., LTD

LOSPA Patella Component Model/Catalog Number: 01.10.9XX Software Version: N/A Product Description: Patella

DateMay 31, 2024 (initiated Mar 29, 2024)
StatusOpen, Classified
ReasonDue to unsupported 10 year expiration date.
Root causeLabeling Change Control
ActionOn 03/29/2024, the firm emailed an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that on 7/1/2020 the firm changed the labeling of their polyethylene (PE) and patella inserts from a 5 year shelf life to a 10 year shelf life. This change was made without adequate documentation evidence to support the increase in shelf life. Customers are instructed to: 1. Inform individuals within your organization who need to be aware of this device recall. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at
Quantity35
DistributionUS Nationwide distribution in the states of AL, AZ, CA, MI, PA, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110404EAUM TOTAL KNEE SYSTEM