Recall Z-1890-2024 ยท NHL Stimulator, Electrical, Implanted, For Parkinsonian Symptoms

Boston Scientific Neuromodulation Corporation

Vercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320

DateMay 24, 2024 (initiated Apr 18, 2024)
StatusOpen, Classified
ReasonDeep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
Root causeDevice Design
ActionOn 4/18/24, correction notices were mailed to Physicians/Surgeons, Hospitals, Healthcare Professionals who were asked to do the following: 1) Review any patient reports of undesired sensations indicative of potential device reset behavior during IPG charging and report these observations accordingly to the firm. 2) After firm verifies this device reset behavior has occurred during IPG charging, an in-field IPG firmware update can be scheduled for the patient s device. This will eliminate the possibility of a coincident routine device system check during IPG charging, thus preventing a potenti
Quantity14,977
DistributionWorldwide - US Nationwide distribution including in the states of TX, WA, MT, GA, TN, MA, OR, ID, CA, FL, WI, UT, DE, AZ, PA, DC, LA, KY, VA, MN, NC, OH, NY, MI, CO, IN, MD, NV, AL, IL, AR, SC, OK, MO, MS, NJ, ME, Hi, KS, SD, WV, NM, AK, CT, WY, NE, ND and the countries of Canada, Argentina, Chile,
FDA recordOpen on fda.gov