Recall Z-1786-2024 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Randox Laboratories Ltd.

Evidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drug molecule and metabolites of drugs in human urine Catalog Number: EV4115

DateMay 9, 2024 (initiated Apr 5, 2024)
StatusOpen, Classified
ReasonFoam gasket on the Evidence MultiSTAT chip heater assembly may not perform as expected. Foam that is not functioning as expected may cause light to leak into the Charged Coupled Device (CCD) Camera during imaging. This will affect the testing and may delay in reporting results.
Root causeDevice Design
ActionRandox Laboratories Ltd UK notified the US Distributor Randox Laboratories via email on 4/05/ 24. End-Users notified by email on 4/9/24. Letter (REC736) states reason for recall, health risk and action to take: " Export support files for the last 10 samples run and share with Randox using the link to be provided " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to technical.services@randox.com within five working days. " Affected chip heater assembly replacement to be arranged by Randox Service Engineer Transmission of Field Saf
Quantity40 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220451Evidence MultiSTAT DOA Urine MultiPlex, Evidence MultiSTAT
Randox Laboratories Limited, Oct 5, 2023