Recall Z-1756-2024 ยท JAK System, X-Ray, Tomography, Computed

Philips North America Llc

The Oncology Essentials Package of the Spectral CT imaging system.

DateMay 2, 2024 (initiated Mar 29, 2024)
StatusOpen, Classified
ReasonPotential for third party oncology marking laser to be reflected off the gantry cover surface, which can potentially lead to misdiagnosis and subsequent incorrect treatment.
Root causeUnder Investigation by firm
ActionAn URGENT Medical Device Correction Letter, dated 3/27/24, was mailed to consignees to notify them of this recall. Consignees are instructed to identify if a third-party laser was installed on their system and if the laser line is reflected, duplicated, or thickened. If a third-party laser is installed, before treatment of a patient, reposition or re-image as necessary to verify proper placement of the localization laser marks. Systems may continue to be used in accordance with the intended use and provided use recommendations. Consignees are to complete and return the provided response form b
Quantity12 units
DistributionWorldwide distribution - US Nationwide in the states of AZ, HI, MN, VT, WV and the countries of China, Macao, New Zealand, Portugal, Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K203020Spectral CT
Philips Medical Systems Nederland B.V., Feb 26, 2021