ClearanceBrief
Recall Z-1773-2024 ยท MSD Catheter, Hemodialysis, Implanted

Angiodynamics, Inc.

DURAMAX STACKED TIP 32 CM STRAIGHT VASC-PAK, 1-UP PG- Indicated for use in attaining Long-Term vascular access for Hemodialysis and Apheresis. Product Number: H787103028201

DateMay 3, 2024 (initiated Mar 18, 2024)
StatusOpen, Classified
ReasonTri-Ball Tunneler may contain loosely affixed metal shavings, potential for the metal shaving to embolize, possibly requiring removal via surgical procedure.
Root causeUnder Investigation by firm
ActionAngiodynamics issued Recall Notifications Packages via Federal Express to 1 Domestic Consignee on 2024-03-18 . Letter states reason for recall, health risk and action to take: IMMEDIATELY o Locate the affected product in your Inventory, cease any and all distribution segregate in a secure location pending return. o Forward a copy of this recall notification to any locations to which you have distributed affected product. 2. Complete and return the Reply Verification Tracking Form. " Promptly complete, sign, and return the enclosed Reply Verification Tracking Form (even if you do not have any
Quantity50 units
DistributionVA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101843DURAMAX CHRONIC HEMODIALYSIS CATHETER AND PROCEDURE KIT
AngioDynamics, Inc., Oct 20, 2010