ClearanceBrief
Recall Z-1781-2024 ยท OWB Interventional Fluoroscopic X-Ray System

GE Medical Systems, SCS

GE Healthcare Allia IGS 7 OR, Interventional fluoroscopic x-ray systems

DateMay 6, 2024 (initiated Mar 18, 2024)
StatusOpen, Classified
ReasonGE HealthCare has become aware of the potential that adhesive material might not have been applied to some bolts of certain Innova IGS, Discovery IGS, Interact IGS, Allia IGS and Allia Pulse IGS products. This could result in a potential fall of the Detector/Elevator or X-ray tube of the frontal Gantry assembly.
Root causeProcess control
ActionGE Healthcare issued an UGENT MEDICAL DEVICE CORRECTION notice to its consignees on 03/18/2024 via letter. The notice explained the issue, hazard, and requested the following actions be taken: Every day prior to first use of the device, perform a functional check of the gantry movements and images of your device to look for the following behaviors: 1. Abnormal noise during gantry movement. 2. Image misalignment. If any of the above is observed, DO NOT use the device and call GE HealthCare Service Immediately. Please ensure all potential users in your facility are made aware of this safety n
Quantity2 units
DistributionWorldwide distribution - US Nationwide in the states of CA, FL, GA, IN, MA, NC, OK, SC, TN, TX, VA and the countries of Armenia, China, France, Hungary, India, Italy, Japan, Kazakhstan, Lithuania, Mexico, Russia, Spain, Taiwan, Turkey, Vietnam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232344Allia IGS 3, Allia IGS 5, Allia IGS 7, Allia IGS 7 OR
GE Medical Systems SCS, Oct 11, 2023