Recall Z-1620-2024 ยท NKB Thoracolumbosacral Pedicle Screw System

SEASPINE ORTHOPEDICS CORPORATION

Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

DateApr 23, 2024 (initiated Feb 28, 2024)
StatusOpen, Classified
ReasonScrew heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
Root causeNonconforming Material/Component
ActionOn 2/26/24, the firm phoned customers and informed them of the recall and that they would receive an email outlining details of the field action and subsequent steps to be taken. On 2/28/24, recall emails were sent to customers asking customers to respond to the emails to coordinate product return and replacement. Customers with questions can contact the firm at olivia.disanto@seaspine.com.
Quantity444 Screw Heads
DistributionUS: CA, FL, CO, TX