Recall Z-1874-2024 ยท DYB Introducer, Catheter

Merit Medical Systems, Inc.

Prelude Guide Sheath Roadster, REF PG5F45S018, 5F 0.076 (1.9 mm), 45 cm, STERILE EO, RX ONLY

DateMay 20, 2024 (initiated Mar 14, 2024)
StatusOpen, Classified
ReasonTheir is a potential that sheaths labeled with .018 dilators may contain a .038 dilator.
Root causeEmployee error
ActionOn March 18, 2024 Merit Medical issued a "Urgent Medical Device Recall Notice to affected consignees via E-Mail. Merit Medical asked consignees to take the following actions: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is im
Quantity18 units
DistributionWorldwide - US Nationwide distribution in the state of CA and the countries of New Zealand and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211405Prelude Guide Sheath
Merit Medical Systems, Inc., Oct 4, 2021