Recall Z-1603-2024 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

IBA Proton Therapy System - PROTEUS 235, PTS-10 versions, PTS-11 versions, and PTS-12 versions.

DateApr 19, 2024 (initiated Mar 4, 2024)
StatusOpen, Classified
ReasonThere is a risk of mistreatment as irradiation is not prevented when some safety parameters are out of tolerance.
Root causeDevice Design
ActionOn 3/20/2024 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers in person or via email by their customer representative. Customers were instructed to review the notice with their customer representative. Immediate actions taken: IBA carried out an analysis to determine if the impacted sites have some safety parameters checked by the BAPP in their clinical site configuration with values leading to violations. The result of this analysis showed that there are no sites with safety parameters checked by the BAPP out of tolerance. Intermediate action: While waiting for the solution to
Quantity8 units
DistributionUS Nationwide distribution in the states of AR, FL, LA, KS, MI, TX, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163500IBA Proton Therapy System- Proteus 235 (Proteus One, Proteus Plus, Proteus TK2, Proteus Class)
Ion Beam Applications S.A., Aug 17, 2017