Recall Z-1511-2024 ยท DXN System, Measurement, Blood-Pressure, Non-Invasive

Baxter Healthcare Corporation

Power Cords, marketed under the following models: a) CORD,METAL EARTH PIN, 2.5A,250V,2.5M,C7, 714682; b) PLFM CLASS II POWER CORD - AUSTRALIA, 716010; c) PLFM CLASS II POWER CORD - EUROPE, 716012; d) POWER CORD,AUSTRALIA,2.5A,250V,2.5M,C7, 008-0910-00; e) POWER CORD,EUROPEAN,2.5A,250V,2.5M,C7, 008-0912-00; f) LINE CORD (EURO), 4500-402; g) LINE CORD (UK), 4500-404; h) LINE CORD (AUS), 45

DateApr 10, 2024 (initiated Mar 26, 2024)
StatusOpen, Classified
ReasonBaxter received reports of an issue related to the construction of the power cord not meeting the insulation rating per country-specific requirements and international electrical standards.
Root causeNonconforming Material/Component
ActionBaxter issued an Urgent Medical Device Correction notice to its consignees on 03/26/2024 via USPS, first class mail. The notice explained the issue, hazard, and provided the following: Actions to be taken by customers: 1. Healthcare providers may continue to use the non-compliant power cords, but if any fraying or damage is observed, users should NOT make physical contact with the cord and discard the power cord immediately. 2. Healthcare providers should regularly inspect the power cords for fraying or other damage. 3. Once Baxter has replacement power cords, a follow-up notification will
Quantity1338 cords
DistributionGlobal, US: NY, IL
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.