Recall Z-1619-2024 ยท LOF Stimulator, Bone Growth, Non-Invasive
EBI, LLC
48" Lead Wires, Replacement Part Number: 1067724-4
| Date | Apr 23, 2024 (initiated Mar 22, 2024) |
| Status | Open, Classified |
| Reason | Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment |
| Root cause | Device Design |
| Action | ZimVie issued Urgent Medical Device Correction letter to Patients, Physicians, Sales Representatives on 3/22/24 . Letter states reason for recall, health risk and action to take: 1. Review this notification for awareness of the contents. 2. Before each use, inspect your device s lead wires (black cable) for damage. If found to be damaged, contact Customer Service at 1-800-526-2579 x 6000 for a complimentary replacement. 3. Complete the Acknowledgement and Receipt Form either by scanning the QR code on the postcard included with this letter and completing the online form or mailing back the com |
| Quantity | 594 units (2 lead wires assembly) |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.