Recall Z-1512-2024 ยท BSZ Gas-Machine, Anesthesia

Draeger Medical, Inc.

Drger Perseus A500 Anesthesia Workstation

DateApr 18, 2024 (initiated Mar 20, 2024)
StatusOpen, Classified
ReasonUnexpected shutdown while operating on battery power.
Root causeLabeling design
ActionConsignees were sent an Urgent Medical Device Recall notification by mail on 3/20/24. The notice informs consignees of the possibility of device shutdown with possible loss of ventilation, which consignees are instructed to either use an emergency ventilation bag or apply manual ventilation with the Perseus with the provided instructions. Consignees are to perform a battery test on all affected devices in their inventory per the provided instructions in the recall notification. If the tested unit(s) fails the battery test, consignees are to contact Drager Service Technical Support between 8:00
Quantity16,841 units
DistributionDomestic: Nationwide Distribution; Foreign: Angola, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Botswana, Brazil, Brunei Dar-se- S, Bulgaria, Canada, Chili, China, Columbia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, French Polynesia, French
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133886PERSEUS A500
Draeger Medical GmbH, Jun 20, 2014