ClearanceBrief
Recall Z-1501-2024 ยท CBK Ventilator, Continuous, Facility Use

Philips Respironics, Inc.

LifeVentEVO2, all software versions except 1.05.06.00. Ventilator for pediatric and adult patients. (sold outside US)

DateApr 17, 2024 (initiated Mar 19, 2024)
StatusOpen, Classified
ReasonPotential for sudden loss of ventilation due to Battery Depleted or Loss of Power alarm while unit has sufficient power.
Root causeDevice Design
ActionConsignees were mailed an URGENT MEDICAL DEVICE CORRECTION notice, dated 3/7/24. Consignees are asked to identify any affected product in their inventory and follow the provided instructions in order to continue using devices safely in CPAP or PSV mode. Devices that alarm are to be immediately plugged into a power source if a Loss of Power Alarm occurs. If no source of power is available, remove and replace the detachable battery. Devices running on software other than 1.05.06.00 can be updated to the latest software version, which does not experience this failure mode. The provided recall not
Quantity1,002 units
DistributionDomestic: Nationwide Distribution; Foreign: AE, AR, AT, AU, BD, BE, BG, BO, BR, BS, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, EC, EE, EG, ES, FI, FR, GB, GE, GF, GI, GR, GT, HK, HR, HU, ID, IE, IL, IN, IT, JE, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, NZ, OM, PA, PE, PF,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181166Trilogy Evo
Respironics, Inc., Jul 18, 2019