Recall Z-1508-2024 ยท JJH Clinical Sample Concentrator

Qiagen Sciences LLC

EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model Number: 9003230

DateApr 10, 2024 (initiated Mar 4, 2024)
StatusOpen, Classified
ReasonInsufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost
Root causeUnder Investigation by firm
ActionQiagen issued Urgent Medical Device Correction letter to Lab Director/Manager, Medical Director, Risk Manager, Safety Officer on 3/4/24 via email. Letter states reason for recall, health risk and action to take: Complete the Acknowledgment of Receipt promptly and return to quality.communications@qiagen.com. Should the error described above occurs, please perform the following actions: o Switch off the instrument and disconnect from the power supply. o Remove any onboard samples and perform decontamination procedure of the instruments as described in the user manual. o Call your local QIAGEN Te
Quantity168 units: 38 units US; 130 units OUS
DistributionNationwide Foreign: AT, AU, BE, CH, CR, CZ, DE, DK, ES, FR, GB, HU, IT, JO, NL, NO, QA, SA, SE, SI, TH
FDA recordOpen on fda.gov