Recall Z-1602-2024 ยท OOI Real Time Nucleic Acid Amplification System

Abbott Molecular, Inc.

Alinity m System, Part Number: 08N53-002

DateApr 19, 2024 (initiated Mar 20, 2024)
StatusOpen, Classified
ReasonThe Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes cou
Root causeUnder Investigation by firm
ActionAbbott notified consignees on about 03/20/2024 via FedEx letter. Consignees were instructed to ensure the sample tube label requirements. If error code 6013 occurs and uncapped sample tubes are present on the sample rack, one of the following options can be used to resolve the error: 1. Aliquot the sample into another tube permitted by the assay package insert and ensure the sample tube label requirements in Appendix A are met; or 2. Cover colored portion of tube label that is present in the Keep-Out Zone with white opaque material (e.g. white sticker) where the barcode is located. Avoid cov
Quantity1,012 units
DistributionWorldwide distribution - US Nationwide and the countries ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CAMBODIA, CAMEROON, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, EL SALVADOR, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERM
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.